Highly selective and cost-effective impedimetric biosensor for cylindrospermopsin detection using an aptamer-functionalized pencil graphite electrode

Küçük Resim Yok

Tarih

2026

Dergi Başlığı

Dergi ISSN

Cilt Başlığı

Yayıncı

Pergamon-Elsevier Science Ltd

Erişim Hakkı

info:eu-repo/semantics/openAccess

Özet

Nutrient runoff from anthropogenic eutrophication, rising atmospheric CO2 levels, and global warming have significantly contributed to the proliferation of harmful cyanobacterial blooms in freshwater ecosystems. These blooms release cyanotoxins, such as cylindrospermopsin (CYN), which pose genotoxic and cytotoxic effects. The detection and monitoring of CYN in water are essential for protecting public health and reducing environmental risks. In this study, an electrochemical aptasensor was developed by effective immobilization of CYN-specific aptamer (cynApt) onto a highly conductive pencil graphite electrode (PGE), which offers a sustainable alternative to expensive gold and graphene-based sensors. CYN was detected through electrochemical impedance spectroscopy (EIS) by monitoring changes in charge transfer resistance. The detection limits for CYN using PGE/ cynApt electrodes were found to be 0.78 ng mL- 1 in deionized water and 0.97 ng mL- 1 in lake water, within a linear range of 1-20 ng mL- 1. The selectivity of the biosensor for CYN was also tested against other cyanotoxins, such as okadaic acid and saxitoxin, both in deionized and lake water.

Açıklama

Anahtar Kelimeler

Cylindrospermopsin, Pencil graphite electrode, Aptamer, Electrochemical biosensor, Impedimetric aptasensor

Kaynak

Toxicon

WoS Q Değeri

Q3